User manual

REMOTE CONTROL
39
Command List
Detailed descriptions of each command start from the next page.
The “HELP” command shows all the below commands and their
meanings, except for the HELP command itself.
ISET<X>:<NR2>
Sets the output current.
ISET<X>?
Returns the output current setting.
VSET<X>:<NR2>
Sets the output voltage.
VSET<X>?
Returns the output voltage setting.
IOUT<X>?
Returns the actual output current.
VOUT<X>?
Returns the actual output voltage.
TRACK<NR1>
Selects the operation mode.
BEEP<BOOLEAN>
Turn on or off the beep.
OUT<BOOLEAN>
Turn on or off the output.
STATUS?
Returns the GPS-3303D or GPS-3303S status.
*IDN?
Returns the GPS-3303D or GPD-3303S
identification.
RCL<NR1>
Recalls a panel setting.
SAV<NR1>
Saves the panel setting.
HELP?
Shows the command list.
ERR?
Returns the instrument error messages.
GPD-3303 Series User Manual
40
Command Details
ISET<X>:<NR2>
Description
Sets the output current.
Panel operation
See page23
Response time
Minimum 70ms
Example ISET1:2.234
ISET1:2.23
Sets the CH1 output current to
2.234A (for GPD-3303S)
Sets the CH1 output current to
2.23A (for GPD-3303D)
ISET<X>?
Description
Returns the output current setting.
Panel operation
See page23
Response time
Minimum 80ms
Example ISET1?
Returns the CH1 output current setting
VSET<X>:<NR2>
Description
Sets the output voltage.
Panel operation
See page23
Response time
Minimum 70ms
Example VSET1:20.345
VSET1:20.3
Sets the CH1 voltage to 20.345V
(for GPD-3303S)
Sets the CH1 voltage to 20.3V
(for GPD-3303D)