OPERATION
27
CH1/CH2 Tracking Series Mode
Background
Tracking series operation doubles the Voltage
capacity of the GPD-3303 series by internally
connecting CH1 (Master) and CH2 (Slave) in serial
and combining the output to a single channel. CH1
(Master) controls the combined Voltage output
level.
The following describes two type of configurations
depending on the common ground usage.
Tracking series without common terminal
Connection
Output rating
0 ~ 60V/0 ~ 3A
1. Press the SER/INDEP key
to activate the tracking
series mode. The key LED
turns on.
GPD-3303 Series User Manual
28
2. Connect the load to the front panel terminals,
CH1+ & CH2 (Single supply).
Note: this diagram shows non-European terminals.
3. Press the CH2 key (LED
turns on) and then use the
Current knob to set the CH2
output current to the
maximum level (3.0A).
By default, the Voltage and
Current knob work in the
coarse mode. To activate the
fine mode, press the knob to
turn the FINE LED on.
(Fine control)
Coarse: 0.1V or 0.1A @ rotation click
Fine: the smallest digit @ rotation click
4. Press the CH1 key (LED
turns on) and then use the
Voltage and Current knob
to set the output voltage
and current level.
5. To turn on the output, press
the output key. The key LED
turns on.
6. Refer to the CH1 (Master) meter and indicator
for the output setting level and CV/CC status.