User manual

OPERATION
25
CH3 Independent Mode
Background /
Connection
The CH3 rating is 2.5V/3.3V/5V, 3A fixed. It
works independently from CH1 and CH2,
regardless of their modes.
Output rating
2.5V/3.3V/5V, 3A fixed
No Tracking
Series/Parallel
CH3 does not have tracking series/parallel mode.
Also, CH3 output is not affected by CH1 and CH2
modes.
Panel operation
1. Connect the load to the front
panel CH3 +/ terminal.
(the diagram shows non-
European terminals)
2. Select the output voltage,
2.5V/3.3V/5V using the
CH3 voltage selector key.
GPD-3303 Series User Manual
26
3. To turn on the output, press
the output key. The key
LED turns on.
CV CC
When the output Current level
exceeds 3A, the overload
indicator turns red and CH3
operation mode switches from
Constant Voltage to Constant
Current.
Note: “overload” in this case does not mean an
abnormal operation.