User Manual

COMMAND OVERVIEW
135
TEMPerature:CONVersion:DISP
Set
Query
Sets or returns the temperature display mode for the
temperature conversion function.
TEMPerature:CONVersion:DISP <NR1>
TEMPerature:CONVersion:DISP?
<NR1>
1:ΔT
2:T
TEMP:CONV:DISP 1
Sets the temperature display mode for the temperature
conversion function is ΔT.
TEMPerature:CONVersion:MATH:DATa
Query
Returns conversion function deviation value.
TEMPerature:CONVersion:MATH:DATa?
<NR3>
±0.000~9.999E±X
TEMP:CONV:MATH:DAT?
Returns 1.250E+2.
TEMPerature:CONVersion:RESistance
Set
Query
Sets or returns the initial resistance for the temperature
conversion function.
TEMPerature:CONVersion:RESistance {<NRf>[,<String>]}
TEMPerature:CONVersion:RESistance?
<NRf>
000.0001~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
<NR3>
000.0001~999.9999E±X