    HANDLER/SCAN INTERFACE 
    93 
Configure GPIB Interface 
Background 
The GPIB interface is SCPI-1994, IEEE488.1 and 
IEEE488.2 compliant. 
  Note 
The GPIB interface is only available on the GOM-804G 
and GOM-805. 
1. Select the 
GPIB address 
Configure the interface to GPIB and set 
the GPIB address in 
System>Utility>Interface menu. 
Page 71 
Connect one end of  the 
GPIB cable to the computer 
and the other end to the 
GPIB port on the 
GOM-805.   
GPIB
RS232/USB Function Check 
Operation 
Invoke a terminal application such as Realterm. 
For RS-232, set the COM port, baud rate, stop bit, data 
bit and parity accordingly.   
To check the COM settings in Windows, see the Device 
Manager. For example, in WinXP go to the Control panel 
→  System  →  Hardware tab. 
Run this query from the terminal. 
*idn? 
This should return the Manufacturer, Model number, and 
Firmware version. 
GWINSTEK,GOM805,GXXXXXXXX,V1.00   
  Note 
If  you are not familiar with using a terminal application 
to send/receive remote commands from the serial port or 
via a USB connection, please page 94 (Using Realterm to 
Establish a Remote Connection) for more information.   