User Manual
    HANDLER/SCAN INTERFACE 
    85 
the measured value from the reference value as a 
percentage.   
{ [(Measured Value-Reference)/Reference]%}. 
A measured value that falls within the upper and lower 
limits is considered IN (pass), a value that falls below the 
lower limits is considered LO, and a value that falls over 
the upper limit is a HI. 
Reference 
value
Upper 
limit
Lower 
limit
IN HILO
For both scan modes, the IN, HI or LO will be shown on 
the display for each judgment (if the time between each 
judgment is not too fast). 
Display Overview 
Scan function 
indicator
Ready to start scan message
Reference, limits, scan mode, current 
channel, measurement delay 
Change 
display view
1. Select the 
Scan function 
Press 
Scan
Scan to access the scan mode, as shown 
above. 
2. Select the 
compare mode 
Use the arrow keys to navigate to the Mode setting. Press 
the Enter key to toggle the compare mode. 
Mode
Move
Toggle
Enter
Range 
Abs,  △ % 