User Manual
    GOM-804 & GOM-805 User Manual 
74 
*The OFF operator sets the Logic as 
true when Operand1 is true. 
** Bin O is defined as outside bin 1~ 8. 
  Note 
The Bin logic settings are not available for the GOM-804. 
Pressing ESC before pressing ENTER will exit from the 
selected External I/O setting. 
Handler Mode 
Background 
The Handler Mode setting determines the behavior of  
the result signals from the handler interface. There are 
two settings, Clear and Hold. The Clear setting will clear 
the results of  the previous test before starting the 
succeeding one and the Hold setting will keep the test 
result of the previous test until the succeeding test has 
completed. 
The timing diagrams below are used as examples. All the 
result signals in the examples are active high. 
Clear example   
Clear: All result signals (PASS, Fail, High and Low) are 
cleared at the falling edge of  EOT and the results from 
the current test are output at the rising edge of  the EOT 
signal. 
Previous 
results
cleared
New 
results
Trigger
Ready
EOT
Pass
Fail
High
Low
EOT falling 
edge
EOT rising 
edge
Previous 
results