User Manual
    MEASUREMENT 
    71 
Interface 
Background 
The remote interface can be set to RS232, GPIB or USB.   
  Note 
The GPIB interface is only available on the GOM-804G 
and the GOM-805. 
1. Select 
Interface setting 
From one of  the main screens, press 
the 
ESC
  key so that the menu 
system at the bottom of  the display 
has focus. 
Go to System and press Enter. 
Go to Utility and press Enter. 
Go to Interface and press Enter. 
System 
menu icon
Move
Select menu 
or setting
Enter
2. Interface 
setting appears 
Use the arrow keys to choose an interface and to set the 
baud rate (RS232) or primary address (GPIB). The EOL 
(end of line) character can also be set. Press Enter to 
confirm the settings. 
       Interface
Interface 
GPIB, Primary Address (1 ~ 30) 
RS232, Baud Rate (1200, 2400, 4800, 
9600, 19200, 38400, 57600, 115200) 
USB 
EOL 
LF, CR, CR+LF, LF+CR (default = LF) 
See page 104 for further details. 
  Note 
Pressing ESC before pressing ENTER will exit from the 
Interface settings. 