User Manual
    MEASUREMENT 
    41 
Compare Function 
Background 
The compare function compares a measured value to a 
“Reference” value that has an upper (HI) and lower (LO) 
limit. If  the measured value is within the upper and lower 
limit, then the measured value is judged as IN.     
There are three compare modes that can be used to make 
a judgment: ABS,  △% and % modes.   
The ABS mode displays the absolute difference between 
the measured and the reference value (shown as  △) and 
compares the measured value to the upper (HI) and 
lower (LO) limit. The upper and lower limits are set as 
absolute resistance values.   
Compare function 
indicator
Absolute 
difference of 
measured value 
from Reference 
value
Measured value
Reference, limits, compare 
mode and beep mode
Pass/Fail 
judgment
A measured value that falls within the upper and lower 
limits is considered IN (pass), a value that falls below the 
lower limits is considered LO, and a value that falls over 
the upper limit is a HI. 
Reference 
value
Upper 
limit
Lower 
limit
IN HILO
[Note that the reference value in the ABS mode is only 
for reference purposes and is not used to make a 
judgment.] 