User Manual
    GOM-804 & GOM-805 User Manual 
16 
GND Terminal 
GND
Connect the GND (ground) terminal to 
the earth ground. 
GUARD Terminal 
GUARD
The GUARD terminal has the same 
potential as earth, but cannot be 
substituted for it. Connect the GUARD 
terminal to the cable shield layer of  the 
test leads to help reduce noise.   
Function Keys 
Ohm
The Ohm key activates the resistance 
measurement function. 
Compare
The Compare key activates the 
comparator function. 
Binning
The Binning key activates the binning 
function to grade the DUTs into eight 
bins according to the tolerance settings. 
GOM-805 only. 
TC
The TC key activates the TC 
(temperature compensation) function 
which calculates the resistance of  a 
DUT at a specified temperature given 
the resistance of  the DUT at the 
ambient temperature and the 
temperature coefficient of the DUT is 
known. 
TCONV
The TCONV (Temperature 
Conversion) function calculates the 
temperature of a DUT given an initial 
temperature, initial resistance, measured 
resistance and a constant (inferred zero 
resistance temperature) for the DUT. 
TEMP
The TEMP key activates the 
temperature measurement function. 