User Manual
    GOM-804 & GOM-805 User Manual 
138 
Trigger Commands 
READ 
Query
Description 
Returns the measurement value. 
Query Syntax 
READ? 
Return parameter 
<NR3> 
±0.0000~5.1000E±X 
Example 
READ? 
>+2.2012E+0 
Returns the measurement. 
MEASure<X> 
Query
Description 
Returns the results of  the selected channel in the scan 
mode, including HI/LO/IN and value. 
Query Syntax 
MEASure<X>? 
Parameter   
<X> 
Channel 1~100 
Return parameter 
0|1|2,<NR3> 
0:LO 
1:IN 
2:HI 
<NR3>: Measurement result. 
Example 
MEAS1? 
>1,+0.9978E+1 
Returns channel 1 as 9.978Ω. 
SHOW 
Query
Description 
Returns the judgments of  all (up to 100) channels in the 
scan mode. 
Query Syntax 
SHOW? 
Return parameter 
<String> 
100 characters 
0:LO 
1:IN 
2:HI 
_:Channel not active   