User Manual
    GOM-804 & GOM-805 User Manual 
136 
Example 
TEMP:CONV:RES 10.00,maohm 
Sets initial resistance value to 10.00MΩ. 
TEMP:CONV:RES? 
>10.0000E+6 
Returns the initial resistance as 10.00MΩ. 
TEMPerature:CONVersion:TEMPerature 
Set
Query
Description 
Sets or returns the initial temperature for the temperature 
conversion function. 
Syntax 
Query Syntax 
TEMPerature:CONVersion:TEMPerature <NRf> 
TEMPerature:CONVersion:TEMPerature? 
Parameter 
<NRf> 
-50.0~399.9 (Unit: ºC) 
Return parameter 
<NR2> 
-50.0~399.9 (Unit: ºC) 
Example 
TEMP:CONV:TEMP 25.6 
Sets the initial temperature to +25.6ºC. 
TEMPerature:DATa 
Query
Description 
Returns the PT-100 sensor temperature measurement in 
degrees Celsius. 
Query Syntax 
TEMPerature:DATa? 
Return parameter 
<NR3> 
-50.0~399.9 
Example 
TEMP:DAT? 
>0.250E+2 
Returns the temperature as 25ºC. 
TEMPerature:STATe 
Set
Query
Description 
Sets or returns the temperature function state. 
Syntax 
Query Syntax 
TEMPerature:STATe {<NR1>|OFF|ON} 
TEMPerature:STATe? 
Parameter/   
Return parameter 
<NR1> 
0:OFF 
1:ON 
OFF 
Turn the temp function off. 
ON 
Turn the temp function on. 










