User Manual
    COMMAND OVERVIEW 
    135 
TEMPerature:CONVersion:DISP 
Set
Query
Description 
Sets or returns the temperature display mode for the 
temperature conversion function. 
Syntax 
Query Syntax 
TEMPerature:CONVersion:DISP <NR1> 
TEMPerature:CONVersion:DISP? 
Parameter/   
Return parameter 
<NR1> 
1:ΔT 
2:T 
Example 
TEMP:CONV:DISP 1 
Sets the temperature display mode for the temperature 
conversion function is  ΔT. 
TEMPerature:CONVersion:MATH:DATa 
Query
Description 
Returns conversion function deviation value. 
Query Syntax 
TEMPerature:CONVersion:MATH:DATa? 
Return parameter 
<NR3> 
±0.000~9.999E±X 
Example 
TEMP:CONV:MATH:DAT? 
Returns 1.250E+2. 
TEMPerature:CONVersion:RESistance 
Set
Query
Description 
Sets or returns the initial resistance for the temperature 
conversion function. 
Syntax   
Query Syntax 
TEMPerature:CONVersion:RESistance {<NRf>[,<String>]} 
TEMPerature:CONVersion:RESistance? 
Parameter 
<NRf> 
000.0001~999.9999 
<String> 
mohm/ohm/kohm/maohm,unit 
If  the unit is not set, the unit will be 
automatically set by the present range. 
Return parameter 
<NR3> 
000.0001~999.9999E±X 