User Manual
    COMMAND OVERVIEW 
    117 
Example 
CALC:SCAN:CHAN 5 
Sets the channel to 5. 
CALCulate:SCAN:DELay 
Set
Query
Description 
Sets or returns the interval delay for the scan function. 
Syntax 
Query Syntax 
CALCulate:SCAN:DELay <NR1> 
CALCulate:SCAN:DELay? 
Parameter/   
Return parameter 
<NR1> 
400~30000 
Unit:ms 
Example 
CALC:SCAN:DEL 500 
Sets interval delay of the scan to 500ms. 
CALCulate:SCAN:LIMit:LOWer 
Set
Query
Description 
Sets or returns the lower limit value for the scan function. 
Syntax 
Query Syntax 
CALCulate:SCAN:LIMit:LOWer {<NRf>[,<String>]} 
CALCulate:SCAN:LIMit:LOWer? 
Parameter 
<NRf> 
000.0000~999.9999 
<String> 
mohm/ohm/kohm/maohm,unit 
If  unit is not set, the unit will be 
automatically set by the present range. 
Return parameter 
<NR3> 
000.0000~999.9999E±X 
Example 
CALC:SCAN:LIM:LOW 0.123,maohm 
Sets the lower limit value to 0.123MΩ. 
CALC:SCAN:LIM:LOW? 
>0.1230E+6 
Returns the lower limit as 0.123MΩ. 
CALCulate:SCAN:LIMit:MODE 
Set
Query
Description 
Sets or returns the scan function compare mode. 
Syntax 
Query Syntax 
CALCulate:SCAN:LIMit:MODE {ABS|DPER} 
CALCulate:SCAN:LIMit:MODE? 
Parameter/   
Return parameter 
ABS 
The test results are judged from 
absolute values. 