User Manual
    GETTING STARTED 
    11 
Advanced 
Temperature 
Measurements 
The GOM-804/805 has a number of  advanced 
temperature functions that can be used with the optional 
temperature probe, PT-100. 
The temperature compensation function can extrapolate 
what the resistance of  a DUT will be at a desired 
temperature, if  the temperature coefficient of  the DUT 
and the resistance of  the DUT at ambient temperature 
are known.   
The temperature conversion function can be used to 
extrapolate what the temperature rise of a DUT will be at 
specified resistance if  the initial resistance, initial 
temperature and the constant for the DUT are known.       
Drive Signals 
The GOM-805 can select a number of different drive 
signals to suit a number of different measurement 
scenarios, for example the Pulse setting can be used to 
cancel the effects of  thermoelectric EMF on the 
measurement results.     
Dry Circuit 
Testing 
Dry circuit testing allows the GOM-805 to measure the 
contact resistance of  switches and connectors according 
to the DIN IEC 512 and ASTM B539 standards. The 
open circuit voltage will not exceed 20mV in this mode 
to prevent the oxidization layer on metal switches and 
connector points from breakdown. GOM-805 only. 
Automatic 
Testing 
For automatic testing The GOM-804/805 has a handler 
interface designed for automatic testing. The handler 
interface outputs the status of PASS, FAIL, HI, LO, 
READY and EOT signals and inputs a trigger control 
signal. Automatic testing is used with the binning, 
compare and scan functions. 
For computer control applications, RS-232 and USB are 
standard remote interfaces, with GPIB as standard only 
for the GOM-805 and GOM-804G.   