User manual

GHT-118 HV/GB Adapter Box
3
1. Connect the red wire to
the High Voltage terminal
of the GPT-9000.
HIGH VOLTAGE
Terminal
2. Connect the white wire to
the Return terminal of the
GPT-9000.
RETURN
Terminal
3. Connect the black wire to
the Source L terminal of
the GPT-9000.
SOURCE L
4. Connect the green/yellow
wire to the GND terminal
at the rear panel of the
GPT-9000.
Ground wire
GND
Internal connection
Internal
connection
diagram
The AC plug L and N
terminals are shorted and
connected to the red HV
wire.
The AC plug GND
terminal is connected to
the white Return wire and
black Source L wire.
The adapters case is
connected to the
green/yellow GND wire.
HV
Return
L N
GND
Case
GND
Source L