User Manual

COMMAND DETAILS
79
:TRIGger:ALTernate
Set
Query
Description
Sets alternating between source triggers on or off.
Syntax :TRIGger:ALTernate {OFF | ON |?}
Parameter OFF Alternate off
ON Alternate on
Return parameter
Returns the Alternate trigger status (ON, OFF).
Example :TRIGger:ALTernate ON
Turns on alternating between source triggers.
:TRIGger:LEVel
Set
Query
Description
Sets or queries the trigger level.
Syntax :TRIGger:LEVel {<NRf> | ?}
Parameter
<NRf> Trigger level value
Return parameter
Returns the trigger level as <NR3>.
Example :TRIGger:LEVel 3.30E-3
Sets the trigger level to time to 330mV/mA.
Note: This command is equal to :TRIGger:HLEVel
System Commands
:SYSTem:LOCK {OFF|ON|?} ...................................... 79
:SYSTem:LOCK {OFF|ON|?}
Set
Query
Description
Turns the panel lock on off.
Syntax :SYSTem:LOCK {OFF|ON|?}