GDS-3000 Programming Manual
30
Return parameter Returns the probe type.
Example :CHANnel1:PROBe:TYPe VOLTage
Sets the Channel 1 probe type to voltage.
:CHANnel<x>:SCALe
Set
Query
Description
Sets or returns the vertical scale. The scale depends
on the probe attenuation factor.
Note the probe attenuation factor should be set
before the scale.
Syntax :CHANnel<x>:SCALe { <NRf> | ?}
Parameter <x>
Channel
<NRf>
Vertical scale
1/2/3/4
CH1/2/3/4 2e–3 ~ 1e+1 2mV ~ 10V
(Probe x1)
Return parameter
<NR3> Returns the vertical scale in volts or amps.
Example :CHANnel1:SCAle 2.00E–2
Sets the Channel 1 vertical scale to 20mV/div