MEASUREMENT
63
Configure Auto-
Range
Press Vertical Only for vertical
autoranging only.
Press Horizontal Only for horizontal
autoranging only.
Press Horizontal and Vertical for
autoranging on both axes.
Limitation
Auto Range does not work in the following
situation.
Input signal frequency is less than 20Hz
Input signal amplitude is less than 30mV
Run/Stop
Background
By default, the waveform on the display is
constantly updated (Run mode). Freezing the
waveform by stopping signal acquisition (Stop
mode) allows flexible observation and analysis. To
enter Stop mode, two methods are available:
pressing the Run/Stop key or using the Single
Trigger mode.
Stop mode icon
When in Stop mode, the Stop icon
appears at the top of the display.
Triggered icon
Freeze waveform
by Run/Stop key
Press the Run/Stop key once.
The waveform and signal
acquisition freezes. To
unfreeze, press the Run/Stop
key again.