User Manual
GDS-3000 Series User Manual
30
First Time Use
Background
This section describes how to connect a signal,
adjust the scale, and compensate the probe. Before
operating the GDS-3000 in a new environment,
run these steps to make sure the instrument
performs at its full potential.
1. Power On
Follow the procedures on the previous page.
2. Set the date
and time
Set the date and time.
Page 145
3. Reset system
Reset the system by recalling the
factory settings. Press the Default
Setup key on the front panel. For
details, see page 161.
4. Install optional
software
The optional software packages
(Power Analysis, Serial Bus Decode)
can be activated. If the optional
software has not been purchased, a
time trial demonstration can be
activated.
Page 193
5. Connect probe
Connect the probe to the Channel1 input terminal
and probe compensation signal output (2Vp-p,
1kHz square wave).
Set the probe attenuation to x10 if the probe has
adjustable attenuation.