User Manual
  COMMAND DETAILS 
51 
Note 
Before using this command, select the 
measurement channel. See the example below. 
Example  :MEASure:SOURce1 CH 1 
:MEASure:PDUTy? 
5.000E+01 
Selects Channel 1, and then measures the positive 
duty cycle. 
:MEASure:PERiod 
Query
Description 
Returns the period. 
Syntax :MEASure:PERiod{?}  
Return parameter  Returns the period as <NR3> 
Note 
Before using this command, select the 
measurement channel. See the example below. 
Example  :MEASure:SOURce1 CH 1 
:MEASure:PERiod? 
1.0E-3 
Selects Channel 1, and then measures the period. 
:MEASure:PWIDth 
Query
Description 
Returns the first positive pulse width. 
Syntax :MEASure:PWIDth{?}  
Return parameter  Returns the first positive pulse width as <NR3> 
Note 
Before using this command, select the 
measurement channel. See the example below. 
Example  :MEASure:SOURce1 CH 1 
:MEASure:PWIDth? 
5.0E-6 
Selects Channel 1, and then measures the positive 
pulse width. 