User Manual
    GDS-3000 Programming Manual 
30 
Return parameter  Returns the probe type. 
Example :CHANnel1:PROBe:TYPe VOLTage 
Sets the Channel 1 probe type to voltage. 
:CHANnel<x>:SCALe 
Set
Query
Description 
Sets or returns the vertical scale. The scale depends 
on the probe attenuation factor. 
Note the probe attenuation factor should be set 
before the scale. 
Syntax  :CHANnel<x>:SCALe { <NRf> | ?} 
Parameter  <x> 
Channel 
<NRf> 
Vertical scale 
  1/2/3/4
CH1/2/3/4  2e–3 ~ 1e+1  2mV ~ 10V 
(Probe x1) 
Return parameter 
<NR3>  Returns the vertical scale in volts or amps. 
Example :CHANnel1:SCAle 2.00E–2 
Sets the Channel 1 vertical scale to 20mV/div 
Math Commands 
:MATH:DISP ................................................................ 31
:MATH:TYPe ................................................................ 31
:MATH:DUAL:SOURce<x> .......................................... 31
:MATH:DUAL:OPERator ............................................. 32
:MATH:DUAL:POSition ............................................... 32
:MATH:DUAL:SCALe ................................................... 33
:MATH:FFT:SOURce ................................................... 33
:MATH:FFT:MAG ........................................................ 34
:MATH:FFT:WINDow .................................................. 34
:MATH:FFT:POSition .................................................. 35
:MATH:FFT:SCALe ...................................................... 35










