User Manual
  COMMAND DETAILS 
27 
Parameter  <x> 
Channel 
<NR3> 
Deskew time 
  1/2/3/4 CH1/2/3/4  -5.00E -
11~5.00E
-11 
-50ns to 50 ns. 
Return parameter 
<NR3>  Returns the deskew time. 
Example :CHANnel1:DESKew 1.300E-9 
Sets the deskew time to 1.3 nano seconds. 
:CHANnel<x>:DISPlay 
Set
Query
Description 
Turns a channel on/off or returns its status. 
Syntax  :CHANnel<x>:DISPlay {OFF | ON | ?} 
Parameter  <x> 
Channel 
Channel on/off 
  1/2/3/4 CH1/2/3/4  OFF Off 
  ON On 
Return Parameter 
ON  Channel is on.  OFF  Channel is off 
Example :CHANnel1:DISPlay ON 
Turns on Channel 1 
:CHANnel<x>:EXPand 
Set
Query
Description 
Sets Expand from ground or from center for a 
channel. Queries the Expand status of a channel. 
Syntax  :CHANnel<x>:EXPand {GND | CENTer | ?} 
Parameter  <x> 
Channel 
GND Ground 
  1/2/3/4 CH1/2/3/4  CENTer Center 
Return parameter 
GND Expand by 
ground 
CENTER  Expand by center 
Example :CHANnel1:EXPand GND 
Sets Channel 1 to Expand by ground. 