User Manual
    GDS-3000 Series User Manual 
200 
Common 
Vertical  Resolution  8 bit 
@1M: 2mV~5V 
@50/75M: 2mV~1V 
 Input Coupling
AC, DC, GND 
  Input 
Impedance 
1M// 15pF 
 DC Gain 
Accuracy 
±(3% X |Readout| + 0.1div + 1mV) 
  Polarity  Normal & Invert 
 Maximum 
Input Voltage 
@1 M: 300V (DC+AC Peak), CAT I  
@50/75: 5 VRMS max 
Offset Position 
Range 
2mV/div ~ 100mV/div : ±0.5V 
200mV/div ~ 5V/div : ±25V 
 Bandwidth 
Limit 
Dependent on the oscilloscope bandwidth (BW. 
BW=150: Full, 20MHz 
BW=250: Full, 20MHz, 100MHz 
BW=350: Full, 20MHz, 100MHz, 200MHz 
  Waveform 
Signal Process 
Add, subtract, multiply, and divide waveforms, FFT, 
FFTrms 
    FFT:Spectral magnitude. Set FFT Vertical Scale to 
Linear RMS or dBV RMS, and FFT Window to 
Rectangular, Hamming, Hanning, or Blackman-
Harris. 
Trigger  Sources  CH1 ,CH2,Line ,EXT 
  Modes  Auto (supports Roll Mode for 100 ms/div and 
slower), Normal, Single 
  Type  Edge, Pulse Width, Video, Pulse Runt, Rise & Fall, 
Alternate, Event-Delay(1~65535 events), Time-
Delay(10nS~10S), I
2
C*, SPI*, UART* 
*optional 
    Runt:Trigger on a pulse that crosses one threshold 
but fails to cross a second threshold before 
crossing the first again. 
    SPI (optional):Trigger on SS, MOSI, MISO, or 
MOSI and MISO on SPI buses. 
 I
2
C (optional):Trigger on Start, Repeated Start, 
Stop, Missing ACK, Address (7 or 10 bit), Data, or 
Address and Data on I
2
C buses. 
 UART (optional): 
Trigger on Tx Start Bit, Rx Start Bit, Tx End of 
Packet, Rx End of Packet, Tx Data, Rx Data, Tx 
Parity Error, and Rx Parity. 