User Manual
    GDS-3000 Series User Manual 
140 
4.  Press Source from the lower menu. 
5.  Use the side menu to select a 
source. 
 Range 
Channel 1 ~ 4 
6.  Press Polarity to toggle the polarity. 
 Range 
Rising edge, falling edge, either. 
7.  Press When from the lower menu. 
8.  Then use the side menu to select 
the condition and width. 
Condition 
> , < , = , ≠ 
Width 
4ns ~ 10s 
9.  Press Threshold from the lower 
bezel to edit the threshold for each 
input source. 
10. Use the side menu to set the upper 
threshold. 
 Range 
-80V~80V 
Set to TTL 1.4V 
Set to ECL -1.3V 
Set to 50% 