User Manual
    GDS-3000 Series User Manual 
128 
Trigger Parameter Overview 
All the following parameters are common for all 
the trigger types unless stated otherwise. 
Trigger source 
CH1 ~ 4 
Channel 1 ~ 4 input signals 
EXT 
External trigger input 
signal 
 AC Line 
AC mains signal 
 Alternate 
Alternate between channel sources for 
the trigger source. 
 EXT Probe 
Probe trigger source. Set the probe as 
either current or voltage. 
Trigger mode 
Auto (un-
triggered 
roll) 
The GDS-3000 generates an internal 
trigger if there is no trigger event, to 
make sure waveforms are constantly 
updated regardless of trigger events. 
Select this mode especially when 
viewing rolling waveforms at slower 
timebases. 
 Normal 
The GDS-3000 acquires a waveform 
only when a trigger event occurs. 
 Single 
The GDS-3000 acquires a 
waveform once when a 
trigger event occurs, then 
stops acquiring. Press the 
Single key to acquire a 
waveform again. 
Coupling DC 
DC coupling. 