GDS-200 & GDS-300 Series User Manual
46
Select the Trigger Source and the Trigger Level
Background
The touch screen can be used to conveniently select
the trigger source as well as set the trigger level.
Select Trigger
Source
Tap on the trigger level indicator to toggle between
trigger sources.
Toggle trigger
source
Set the Trigger
Level
Swipe the screen to the right of the grid to set the
trigger level with the currently selected source.
The trigger level will also be shown in a popup
window.
Example
Trigger
Level