Datasheet

Figure 1: Graph of Capturing Burst Signal of Traditional Oscilloscope
As can be seen from figure 2, the segmented memory function take
advantages of the memory space with efficiency. The burst signal will be
split and captured into different memory sections under the same
trigger condition. It is about maintaining the high sampling rate and
sustaining the original integrity of waveforms.
Figure 2: Graph of Segmented Memory Function
How to execute the segmented memory function on GDS-2000A
oscilloscope?
Please refer to figure 3, first of all, select the Acquire” bottom at
upper-right hand side, then select the “Segments” button at the bottom
of the screen, then subsequently turn on the Segments ONFunction.
After choosing the triggering condition and the measure parameters,
then set the number of segment sections, which can be divided into
2048 as the maximum, at the upper-right hand side of the screen.
Finally press the “Segment RUN”, then when finish capturing the
waveforms, press “Segment STOP. Choosing the “Current Seg at the
upper-right hand side of the screen, the burst signal in each memory
segment can be easily observed. As can be seen in Figure 4, due to the