User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Utility Commands
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Go_NoGo Commands
- Replay Commands
- Rotate Commands
- DMM Commands
- APPENDX
- INDEX
COMMAND DETAILS
99
Return parameter
Returns the trigger mode.
Example
:TRIGger:MODe NORMal
Sets the trigger mode to normal.
:TRIGger:HOLDoff
Description
Sets or queries the holdoff time.
Syntax
:TRIGger:HOLDoff {<NRf> | ?}
Parameter
<NRf>
Holdoff time
Return parameter
<NR3>
Returns the trigger holdoff time.
Example
:TRIGger:HOLDoff 1.00E-8
Sets the trigger holdoff time to 10ns.
:TRIGger:LEVel
Description
Sets or queries the level.
Syntax
:TRIGger:LEVel {TTL | ECL | SETTO50 | <NRf> | ?}
Related
commands
:TRIGger:TYPe
Parameter
<NRf>
Trigger level value
TTL
Sets the trigger level to TTL.
ECL
Sets the trigger level to ECL.
SETTO50
Sets the trigger level to 50%
Return parameter
<NR3>
Returns the trigger level.
Example1
:TRIGger:LEVel TTL
Sets the trigger to TTL.
Example2
:TRIGger:LEVel 3.30E-1
Sets the trigger level to 330mV/mA.
Set
Query
Set