User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Utility Commands
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Go_NoGo Commands
- Replay Commands
- Rotate Commands
- DMM Commands
- APPENDX
- INDEX
GDS-200 & GDS-300 Programming Manual
84
:MEASUrement:MEAS<X>:MEAN
Description
Returns the mean measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:MEAN?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 4.
Example
:MEASUrement:MEAS3:SOUrce1 CH1
:MEASUrement:MEAS3:TYPe PK2PK
:MEASUrement:STATIstics:MODe ON
:MEASUrement:MEAS3:MEAN?
2.090E-02
Returns the mean measurement result for
measurement number 3.
:MEASUrement:MEAS<X>:MINImum
Description
Returns the minimum measurement results for the
selected measurement from the last time the
statistics were reset. This is a statistics related
command.
Syntax
:MEASUrement:MEAS<X>:MINImum?
Related
Commands
:MEASUrement:STATIstics:MODe
Parameter
MEAS<X>
The automatic measurement number
from 1 to 4.
Query
Query