COMMAND DETAILS
45
HV
Turns horizontal and vertical cursors on.
Return parameter
Returns the state of the cursors (H, HV, OFF).
Example
:CURSor:MODe OFF
Turns the cursors off.
:CURSor:SOURce
Description
Sets or queries the cursor source.
Syntax
:CURSor:SOURce {CH1 | CH2 | REF1 | REF2 | MATH |
| ?}
Parameter
CH1~CH2
Channel 1 to 2
REF1~2
Reference waveform 1 to 2
MATH
Math source
Return parameter
Returns the cursor source.
Example
:CURSor:SOURce CH1
Turns the cursor source as channel 1.
:CURSor:HUNI
Description
Sets or queries the units for the horizontal bar
cursors.
Syntax
:CURSor:HUNI {SEConds | HERtz | ?}
Related
Commands
:CURSor:MODe
Parameter
SEConds
Sets the cursor units to time in seconds.
HERtz
Sets the cursor units to frequency.
Return parameter
Returns the unit type.
Example
:CURSor:HUNI SEConds
Sets the units to time in seconds.
Set
Query
Set
Query