User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Utility Commands
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Go_NoGo Commands
- Replay Commands
- Rotate Commands
- DMM Commands
- APPENDX
- INDEX
COMMAND DETAILS
33
Vertical Commands
:CHANnel<X>:BWLimit .............................................. 33
:CHANnel<X>:COUPling ............................................ 33
:CHANnel<X>:DISPlay ............................................... 34
:CHANnel<X>:EXPand ............................................... 34
:CHANnel<X>:INVert ................................................. 35
:CHANnel<X>:POSition ............................................. 35
:CHANnel<X>:PROBe:RATio ..................................... 36
:CHANnel<X>:PROBe:TYPe ....................................... 36
:CHANnel<X>:SCALe ................................................. 36
:CHANnel<X>:BWLimit
Description
Sets or returns the bandwidth limit (20MHz)
on/off.
Syntax
:CHANnel<X>:BWLimit {OFF | ON | ?}
Parameter
<X>
Specified channel 1,2
ON
Bandwidth limit is on.
OFF
Bandwidth limit is off.
Return parameter
Returns the bandwidth limit status for the selected
channel.
Example
:CHANnel1:BWLimit ON
Sets the channel 1 bandwidth limit on.
:CHANnel<X>:COUPling
Description
Selects or returns the coupling mode.
Syntax
CHANnel<X>:COUPling {AC | DC | GND | ?}
Parameter
<X>
Channel 1,2
AC
AC coupling
Set
Query
Set
Query