User manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Utility Commands
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Go_NoGo Commands
- Replay Commands
- Rotate Commands
- DMM Commands
- APPENDX
- INDEX
COMMAND DETAILS
25
HORizontal OFF;TRIGger OFF;MEASUrement OFF;CURSor
OFF;CURSor:MEASure OFF;:BUZZER OFF
*OPC
Description
The *OPC command sets the OPC bit (bit0) of the
Standard Event Status Register when all current
commands have been processed.
The *OPC? Query returns 1 when all the
outstanding commands have completed.
Syntax
*OPC, *OPC?
Return parameter
1
Returns 1 when all the outstanding
commands have completed.
Example
*OPC?
>1
*RCL
Description
Recalls a panel setup.
Syntax
*RCL {1 | 2 | 3 |…. | 20}
Example
*RCL 1
Recalls the setup from Set 1.
*RST
Description
Resets the GDS-200/300 (recalls the default panel
settings).
Syntax
*RST
*SAV
Description
Saves the current panel setup to the selected setup
number file number.
Syntax
*SAV {1 | 2 | 3 |…. | 20}
Set
Query
Set
Set
Set