GDS-200 & GDS-300 Programming Manual
102
Return parameter
<NR3>
Returns the video trigger line.
Example
:TRIGger:VIDeo:LINe 1
Sets the video trigger to line 1.
:TRIGger:VIDeo:POLarity
Description
Sets or queries the video trigger polarity.
Syntax
:TRIGger:VIDeo:POLarity { POSitive | NEGative | ? }
Related
commands
:TRIGger:TYPe
Parameter
POSitive
Positive polarity
NEGative
Negative polarity
Return parameter
Returns the video trigger polarity.
Example
:TRIGger:VIDeo:POLarity POSitive
Sets the video trigger polarity to positive.
:TRIGger:PULSe:WHEn
Description
Sets or queries the pulse width trigger conditions.
Syntax
:TRIGger:PULSe:WHEn { MOREthan | LESSthan |
EQual | UNEQual | ? }
Related
commands
:TRIGger:TYPe
:TRIGger:PULSe:TIMe
Parameter
MORE than
>
LESSthan
<
EQual
=
UNEQual
≠
Return parameter
Returns the pulse width trigger conditions.
Example
:TRIGger:PULSe:WHEn UNEQual
Sets the trigger pulse width conditions to not equal
to (≠).
Set
Query
Set
Query