OPERATION
73
Trigger Level
Background
The trigger level can be set using touch gestures
or from the trigger level.
Touch Control
1. Select the trigger source from the side of the
graticule.
2. Swipe up and down from the right side of the
graticule to set the trigger level for the selected
trigger source.
Menu Control
1. From the Trigger menu use the Trigger Level
slider to set the trigger level.
2. To set the trigger level to 50%, drag from the
Trigger Level button and drop to the “Set to
50%” option.
3. To manually set the trigger level, drag from the
Trigger Level button and drop to the keyboard
icon. Type in the value for the trigger level with
the keypad that appears.
Drag and drop the
Trigger Level
button to
manually set the
trigger level
Trigger Level
button