OPERATION
105
Calibration
Background
The Calibration menu allows you to perform
vertical calibration on the oscilloscope channels.
Steps
1. From the Utility menu, press
Calibration>Vertical Calibration to start vertical
calibration.
2. A prompt will appear asking you to connect the
CAL output to CH1.
3. Connect the Vertical Calibration cable, GCL-
001, to the input in the restricted panel* on the
interface panel and to CH1.
*You may need to use a Philips head
screwdriver to remove the panel.
1
GCL-001,
Vertical
Calibration
cable
Input on
the
restricted
access tab
V
W
COMA mA
FUSED
10A MAX
FUSED
CAT II600mA MAX 600V
CAT III 300V
4. Press OK to start the CH1 calibration.