User Manual
    GDS-1000A-U Series User Manual 
110 
 Detail 
Contains the waveform amplitude and 
time of each point (4k/1M/2M) 
relative to the trigger point. 
 Fast 
Only contains the waveform 
amplitude data for each point 
(4k/1M/2M). 
Waveform type  CH1, 2 
Input channel signal 
 Math 
Math operation result (page 63) 
Storage location  Internal 
memory 
The oscilloscope’s internal memory, 
which can hold 15 waveforms. 
 External 
USB Flash 
drive 
A USB flash drive (FAT or FAT32 
format) can hold practically an 
unlimited number of waveforms. 
 Ref A, B 
Two reference waveforms are used as 
a buffer to recall a waveform in the 
display. You have to save a waveform 
into internal memory or to USB, then 
copy the waveform into the reference 
waveform slot (A or B), and then recall 
the reference waveform into the 
display. 
Waveform 
Memory Depth 
The memory depth is limited to 1 M points when 
both channels are activated or 2M points when 
only a single channel is activated. The signal must 
be triggered /stopped to have access to the full 
memory depth. Therefore when a signal is saved 
the waveform will be automatically stopped if it is 
not manually triggered /stopped first. 
There are a number of conditions when all of the 
available memory is not utilized due to a limited 
number of different sample rates. This can be 
caused by an un-triggered signal, or a time/div 
setting that is too fast to display all the points on 
screen. 