Datasheet
Timeout, Alternate, Event-Delay(1~65535 events),
Time-Delay(Duration, 4nS~10S)
Holdoff range
4ns to 10s
Coupling
AC, DC, LF rej., Hf rej., Noise rej.
Sensitivity
1div
External Trigger
Range
±15V
Sensitivity
DC ~ 100MHz Approx. 100mV
100MHz ~ 200MHz Approx. 150mV
Input Impedance
1MΩ±3%~16pF
Horizontal
Time base Range
5ns/div ~ 100s/div (1-2-5 increments)
ROLL:
100ms/div ~ 100s/div
Pre-trigger
10 div maximum
Post-trigger
2,000,000 div maximum
Timebase Accuracy
±50 ppm over any ≥ 1 ms time interval
Real Time Sample
Rate
1GSa/s max.
Record Length
Max. 10Mpts
Acquisition Mode
Normal, Average, Peak Detect, Single
Peak Detection
2nS (typical)
Average
selectable from 2 to 256
X-Y Mode
X-Axis Input
Channel 1; Channel 3*
*four channel models only
Y-Axis Input
Channel 2; Channel 4*
*four channel models only
Phase Shift
±3° at 100kHz
Cursors and
Measurement
Cursors
Amplitude, Time, Gating available; Unit:
Seconds(s), Hz(1/s), Phase(degree), Ration(%)
Automatic
Measurement
36 sets: Pk-Pk, Max, Min, Amplitude, High, Low,
Mean, Cycle Mean, RMS, Cycle RMS, Area, Cycle
Area, ROVShoot, FOVShoot, RPREShoot,
FPREShoot, Frequency, Period, RiseTime, FallTime,
+Width, -Width, Duty Cycle, +Pulses, -Pulses,
+Edges, -Edges, FRR, FRF, FFR, FFF, LRR, LRF,
LFR, LFF, Phase
Cursors measurement
Voltage difference between cursors ( ΔV) Time
difference between cursors ( ΔT)
Auto counter
6 digits, range from 2Hz minimum to the rated
bandwidth
Control Panel
Function
Autoset
Single-button, automatic setup of all channels for
vertical, horizontal and trigger systems, with undo
Autoset
Save Setup
20set