User Manual Manual
APPENDIX
151
AC Voltage, ACV+DCV[3] (AC Coupled)
Range
Resolution
Full
Scale
Accuracy (1 year 23˚C ±5˚C) [1]
30-50Hz
50-10kHz
10K-30kHz
30K-100kHz
500mV
10μV
510.00
1.00%+40
0.50%+40
2.00%+60
3.00%+120
5V
100μV
5.1000
1.00%+20
0.35%+15
1.00%+20
3.00%+50
50V
1mV
51.000
1.00%+20
0.35%+15
1.00%+20
3.00%+50
500V
10mV
510.00
x
0.5%+15
1.00%+20[2]
3.00%+50[2]
750V
100mV
765.0
x
0.5%+15
x
x
[1]Specifications are for sine wave inputs that are greater than 5% range.
[2]Input voltage <300Vrms.
[3]The accuracy of ACV+DCV is equal to ACV’s with 10 more digits added.
* The specifications are guaranteed to an input of 750V. A beeping alarm will go off
when the input value is higher than 750V.
* Input protection of 1000V peak on all ranges.
* AC-coupled true RMS – measures the AC component of the input with up to
400Vdc of bias on any range.
AC Current, ACI+DCI[3] (AC Coupled)
Range
Resolu-
tion
Full
Scale
Accuracy (1 year 23˚C ±5˚C) [1]
Burden
Voltage
30-50Hz
50-2kHz
2K-5kHz
5K-20kHz
500μA
10nA
510.00
1.50%+50
0.50%+40
1.50%+50
3.00%+75
0.06V max
5mA
100nA
5.1000
1.50%+40
0.50%+20
1.50%+40
3.00%+60
0.6V max
50mA
1μA
51.000
1.50%+40
0.50%+20
1.50%+40
3.00%+60
0.14V max
500mA
10μA
510.00
1.50%+40
0.50%+20
1.50%+40
3.00%+60[2]
1.4V max
5A
100μA
5.1000
2.0%+40
0.50%+30
x
x
0.5V max
10A
1mA
12.000
2.0%+40
0.50%+30
x
x
0.8V max
[1] The 500μA range requires an input of >35μA to meet specifications. The
5mA~10A ranges need more than 5% of full scale range to meet specifications.
[2] Input current (5k ~ 20kHz)<330mArms.
[3]The accuracy of ACI+DCI is equal to ACI’s with 10 more digits added.
* The specifications are guaranteed to 10A. A beeping alarm will go off when the
input current being measured is higher than 10A.