REMOTE CONTROL
131
4. Press the Down or UP keys
repeatedly to select the baud rate.
2304001152005760038400192009600
5. Press the Enter key followed by
the Exit key. The RS-232 setting
is stored and the display goes
back to the default display.
AUTO
ENTER
/ EXITSHIFT
6. Connect the RS-232C cable to
the rear panel terminal.
RS-232C pin
assignment
Pin 2: RxD
Pin 3: TxD
Pin 5: GND
Pin 1, 4, 6 ~ 9: No Connection
5 4 3 2 1
9 8 7 6
PC – GDM
RS-232C
Connection
A null-modem connection, in which transmit (TxD) and
receive (RxD) lines are cross-linked, is required.
GDM PC
TxD
RxD
GND
Pin2
Pin3
Pin5
TxD
RxD
GND
Pin2
Pin3
Pin5
Set the EOL Character
Description
The TX TERM configuration menu can set the
end-of-line (EOL) character for remote commands. The
GPIB EOL character is fixed as CR+LF.
EOL
CR, LF, CR+LF (default = CR+LF)
Configuration
1. Press the Shift key, the 2nd
(Menu) key, the left key
repeatedly until the TX TERM
configuration menu appears.
/ EXITSHIFT
2ND
MENU
HOLD