Manual
GDM-350B User Manual
26
DC Current
Range
Resolution
Accuracy
2000µA
1µA
±(1% Reading +2 Digits)
20mA
0.01mA
±(1% Reading + 2 Digits)
200mA
0.1mA
±(1.2% Reading +2 Digits)
10A
0.01A
±(2% Reading +5 Digits)
Overload Protection:
mA range: F2 fuseφ6×25mm, F 1A H 240V (CE)
10A range: F1 fuseφ6×25mm, F 10A H 240V (CE)
Warning:
When ≤ 5A Continuous measurement is allowed.
When > 5A Measurements must not take longer than 10 seconds with a wait of 15
minutes between measurements.