Remote Sense 
Remote Sense 
Operation 
Remote sense is used to compensate for the 
voltage drop seen across load cables due to 
resistance inherent in the load cables. The 
remote sense function can compensate a 
maximum of 5% of the output voltage and all 
of output frequency. 
1.  Configure the remote sense setting to ON 
(page 121). 
2.  Connect the Neutral terminal of the remote 
sense terminal block to the Neutral terminal of 
the load. 
3.  Connect the Live terminal of the remote sense 
terminal block to the Live terminal of the load. 
Connection 
Example 
OUTPUT 
terminal 
block
Load 
cables
Sensing 
points
Sensing 
cables
N.C. terminals are Not 
available in the case.
Load