User Manual
  OPERATION 
97 
Advanced Settings 
  Using the Remote Sense Function → from page 97 
  Preset Settings → from page 100 
Using the Remote Sense Function 
The ASR-2000 can be operated using local or remote voltage sense. 
By default, the power supply is configured for local sense.  
 WARNING 
Ensure the output is off before handling the 
remote sense connectors. 
Use sense cables with a voltage rating exceeding 
the isolation voltage of the power supply. 
Never connect sensing cables when the output is 
on. Electric shock or damage to the power supply 
could result. 
Remote Sensing 
Input Connectors 
Overview 
The remote sensing input 
connector is located at the rear 
panel of the ASR-2000. 
Local Sense 
Local Sense 
Operation 
When using local sense, the remote sensing 
input terminal is not used. No compensation of 
any possible voltage drop seen on the load 
cables is performed. Local sense is only 
recommended when the voltage drop is of no 
consequence. By default, the power supply is 
configured for local sense. 
1.  Check that the remote sense setting is disabled 
(page 121). 