User Manual
    ASR-2000 Series User Manual 
84 
Setting the Output Off Phase 
Background 
The off phase setting sets the ending phase of the 
voltage output. 
Steps 
1.  Use the scroll wheel followed by 
the Enter key to make the OFF 
Phs parameter selectable. 
2.  Set the OFF Phs setting with the scroll wheel or 
with the F3 (MAX) and F4 (MIN) soft-keys to set 
the Off Phase to the maximum and minimum 
values, respectively. 
AC+DC-INT, AC-INT, AC+DC-ADD, 
AC-ADD, AC+DC-Sync, AC-Sync 
OFF 
Phs 
Range 
0.0° ~ 359.9° 
Soft-keys 
FIXED/FREE, MAX, MIN 
3.  Press Enter to confirm the Off Phase setting. 