User Manual
    ASR-2000 Series User Manual 
38 
Output Terminals 
Background 
The output terminals can be output from either 
the front panel or from the rear panel. The 
outputs are limited to 5 A / 2.5 A (ASR-2050), 
10 A / 5 A (ASR-2100). 
Supported Plugs 
Multi-region terminal socket 
Supported standards 
IEC, North America, Japan. 
EURO CEE type universal plug 
WARNING 
Dangerous voltages. Ensure that the power to the 
instrument is disabled before handling the power 
supply output terminals. Failing to do so may lead 
to electric shock. 
CAUTION 
For the front panel output, the maximum output 
voltage is 250 VAC and current is 10 A. 
Front Panel 
Output 
Connection 
1.  The front panel has a multi-region power 
socket depending on the socket type. 
2.  Insert the plug from the DUT into the socket.  
(ASR-2000 or ASR-
2000R w/t GET 
Series Box) 
EURO CEE socket 
IEC North America, Japan 
 WARNING 
Dangerous voltages. Ensure output is off before 
unplugging the plug from the front panel socket. 










