Error Messages & Messages 
The following error messages or messages may appear on the ASR-
2000 screen display during varied operations. 
Error Messages 
Description 
Protection type 
Over Ipeak+ Current 
Positive output current peak value 
is excessive. Press "Shift + Cancel" 
to clear this alarm. 
Output Off 
Over Ipeak- Current 
Negative output current peak 
value is excessive. Press "Shift + 
Cancel" to clear this alarm. 
Output Off 
Overheat 
Internal power stage over heat. 
Press "Shift + Cancel" to clear this 
alarm. Check if operating 
temperature is suitable and if filter 
is clogged. 
Output Off 
Over Irms Current 
Output current RMS value is 
excessive. Press "Irms" to check 
allowance set range 
Output Off 
Power Input Anomaly 
The power input voltage is 
insufficient or turning off main 
power switch. Check input power 
before rebooting the unit. 
System Lock 
Fan Failure 
Fan failure. Contact service center. 
System Lock 
Output Over-Power 
Over internal power stage 
maximum power (110% of rating 
power), press "Shift + Cancel" to 
clear this alarm. 
Output Off 
Output Short 
Call attention to output terminal 
short status 
Output Off 
Output Overvoltage 
Over internal maximum voltage 
(110% of rating voltage). Press 
"Shift + Cancel" to clear this 
alarm. 
Output Off 
Calibration Data Error 
The calibration data is abnormal 
or out of allowance range. Contact 
service center. 
Output Off 