User Manual
    ASR-2000 Series User Manual 
150 
Time 
0.0001 ~ 999.9999s (Normal1, 
Normal2 and Abnormal)  
0.0000 ~ 999.9999s (Trans1 and 
Trans2) 
Note: For Trans1 and Trans2, it 
supports a value of 0, which will 
skip the step. 
6.  In order to adjust ACV voltage range between 
HI and LO, it is required to set up outside of 
the SIMULATE menu. Refer to the page 62 for 
details. The selected range will be shown on 
the top bar. 
Range
Range 
LO - 100V, HI - 200V 
7.  Go to the ACV setting and set the Vrms level 
of the step. If you input an ACV value that is 
not within the voltage range, the warning 
message below will be shown. 
Not applicable for Trans1, Trans2 and Normal2. 
ACV 
0.0 ~ 350.0V (Range 200V) 
0.0 ~ 175.0V (Range 100V) 