User Manual
  TEST MODE FUNCTION 
131 
AC/DC Voltage 
Range (ACV/DCV) 
There are 2 voltage range settings: 
HI 200V & LO 100V, which result in 
varied ranges of ACV and DCV 
values, respectively. 
Fset (Frequency) 
Sets the frequency of the step. 
There are 3 secondary frequency 
settings that determine how the 
frequency is output. 
CT: Sets the frequency level of the 
step immediately to Fset values.  
KP: Sets the frequency level to 
“keep” the frequency of the 
previous step. 
SP: Linearly increases or decreases 
the frequency from the end of the 
previous step to the end of the 
current step. 
Note: it is available for AC+DC-INT 
and AC-INT modes only. 
Wave 
Sets the outputting waveform of the 
step. Up to 4 waves including sine, 
square, triangle and arbitrary (1-16) 
wave shapes are available. 
Note: it is available for AC+DC-INT 
and AC-INT modes only. 
Jump To 
The Jump To setting determines 
which step to jump to at the end of 
the step. If Jump To is turned off, 
the unit will follow the Term (Step 
termination) setting for the step. 
Jump Cnt 
Determines the number of times to 
loop the jump step. 