User Manual
  MISCELLANEOUS 
121 
Remote Sense 
The remote sense function detects the output voltage at the sensing 
input terminal. This function compensates for voltage drops across 
the load cables when the load is connected to the ASR-2000 over a 
long distance. 
 Note 
The remote sense function can compensate a 
maximum of 5% of the output voltage. The 
maximum output voltage when compensation is 
used is limited by the rated voltage. 
Steps 
1.  Press the Menu key. The Menu 
setting will appear on the display. 
2.  Use the scroll wheel to go to item 2, MISC 
Configuration and press Enter. 
3.  Go to the Remote Sense setting using the scroll 
wheel and press Enter. Turn the setting on or 
off and press Enter again to confirm. 
Remote Sense 
ON, OFF 
Exit 
4.  Press Exit[F4] to exit from the 
MISC Configuration settings. 
Example 