Power ON 
The Power ON setting allows you to have the power-on output or 
other operation functions on automatically after startup. The 
settings that are loaded are the last settings that were present in the 
standard mode before the unit was turned off last. 
Steps 
1.  Press the Menu key. The Menu 
setting will appear on the display. 
2.  Use the scroll wheel to go to item 2, MISC 
Configuration and press Enter. 
3.  Go to the Power ON setting using the scroll 
wheel and press Enter. Select a setting and 
press Enter to confirm. 
ON 
Set power-on output ON with the 
setting that was loaded before the 
unit was last turned off. 
OFF 
Disable this function active. 
SEQ 
Execute the sequence that was 
loaded before the unit was last 
turned off. 
SIM 
Execute the simulation that was 
loaded before the unit was last 
turned off. 
Exit 
4.  Press Exit[F4] to exit from the 
MISC Configuration settings. 